1.Purpose
To sterilize laboratory materials and waste reliably using the autoclave, while protecting operators from steam and pressure hazards and confirming each cycle achieves sterilization.
2.Scope
Applies to steam autoclaves used to sterilize glassware, media, instruments and biohazard waste. Chemical disinfection of surfaces is covered by the lab cleaning SOP.
Definitions
- Gravity cycle
- A standard sterilization cycle relying on gravity displacement of air by steam, used for most solid goods.
- Liquid cycle
- A slower-exhaust cycle used for liquids to prevent boil-over from rapid pressure release.
- Biological indicator
- A test vial containing heat-resistant spores used to confirm a cycle actually achieved sterilization.
3.Responsibilities
- Lab Technician
- Loads, runs and unloads the autoclave, and records cycle parameters and indicator results.
- Laboratory Manager
- Reviews weekly biological indicator results and approves the autoclave's return to service after a failure.
- Maintenance Technician
- Services the autoclave, checks water levels and gaskets, and repairs faults.
RACI matrix
| Activity | Lab Technician | Laboratory Manager | Maintenance Technician |
|---|---|---|---|
| Check water level and door seal | R/A | - | C |
| Select cycle and load items | R/A | - | - |
| Run weekly biological indicator test | R | A | - |
| Approve return to service after a failed load | I | A | R |
R = Responsible, A = Accountable, C = Consulted, I = Informed
4.Materials and PPE
Materials, tools and systems
- โAutoclave-safe bags and containers
- โChemical indicator tape or strips
- โBiological indicator vials and incubator
- โAutoclave cycle log
- โHeat-resistant gloves and tray for unloading
Personal protective equipment
- โHeat-resistant gloves
- โFace shield or safety glasses
- โClosed-toe shoes
5.Procedure
- 5.1
Check the water reservoir
Lab TechnicianCheck the autoclave's water reservoir level and top it up with the manufacturer-specified water type before starting a cycle.
- 5.2
Inspect the door seal
Lab TechnicianInspect the door gasket for cracks, debris or misalignment, since a damaged seal will prevent the chamber from reaching pressure.
- 5.3
Load items correctly
Lab TechnicianPlace items in autoclave-safe containers or bags, venting biohazard bags loosely and avoiding overpacking so steam can circulate around every item.
Warning: Never seal a biohazard bag tightly closed; steam must be able to penetrate the load.
- 5.4
Apply a chemical indicator
Lab TechnicianPlace a chemical indicator strip or tape on or inside each load so a color change confirms exposure to autoclave conditions.
- 5.5
Select the correct cycle
Lab TechnicianChoose the gravity, liquid or waste cycle matching the load type, since liquids and waste need slower exhaust to avoid boil-over.
- 5.6
Close and lock the door
Lab TechnicianClose the chamber door and confirm it is fully locked before starting the cycle, following the unit's specific locking procedure.
- 5.7
Start the cycle and record parameters
Lab TechnicianStart the selected cycle and record the temperature, pressure, run time and operator on the autoclave cycle log.
- 5.8
Monitor for completion
Lab TechnicianCheck the display or printout at the end of the cycle to confirm the sterilization phase reached and held the required temperature and time.
Checkpoint: The cycle printout or display confirms the sterilization set point was reached and held for the full cycle time.
- 5.9
Allow pressure to release before opening
Lab TechnicianWait until the chamber pressure gauge reads zero and any manufacturer-specified cooling time has passed before unlocking the door.
Warning: Never force open the door while the chamber shows any residual pressure.
- 5.10
Unload with heat-resistant gloves
Lab TechnicianOpen the door slowly to let residual steam escape away from your face, then remove items using heat-resistant gloves and a tray.
- 5.11
Verify sterilization results
Lab TechnicianCheck the chemical indicator color change on each item, and run and read a weekly biological indicator test against a positive control.
Checkpoint: The weekly biological indicator shows no growth compared with the positive control before the autoclave continues routine use.
- 5.12
Log the cycle and report failures
Lab TechnicianComplete the cycle log entry with indicator results, and report a failed cycle or failed biological indicator to the laboratory manager immediately, holding the load as non-sterile.
6.Quality checks
- โEvery load carries a chemical indicator that shows the expected color change.
- โCycle logs show the sterilization temperature and time were reached and held.
- โA biological indicator test is run and read at least weekly.
- โFailed loads are held as non-sterile and reprocessed, not released.
7.Records
- โAutoclave cycle log
- โBiological indicator test results
- โChemical indicator records
- โMaintenance and repair history
8.KPIs
- โPercentage of cycles passing biological indicator verification
- โNumber of failed or aborted cycles per month
- โAverage autoclave downtime for repair
- โPercentage of weekly biological indicator tests completed on schedule
9.Common mistakes
- โSealing a biohazard bag tightly so steam cannot reach the contents.
- โOpening the door before the pressure gauge reads zero.
- โOverpacking the chamber so steam cannot circulate around every item.
- โSkipping the weekly biological indicator test when the schedule is busy.
10.Revision history