Lab Equipment Calibration SOP Template

A ready-to-use SOP for calibrating laboratory instruments such as balances, pH meters, ovens and spectrophotometers. Covers scheduling, reference standards, tolerance checks and labeling. Pipettes are covered by a separate SOP.

FreeNo sign-upWord, PDF, Excel & CSV
Ilia PirozhenkoReviewed by Ilia Pirozhenko, Founder, Perfect WikiUpdated September 15, 202612 steps4 roles3 min read
Standard operating procedureSOP-LAB-002 ยท Rev 1.0
Owner
Quality Assurance Manager
Effective date
September 15, 2026
Review cycle
Every 12 months

1.Purpose

To ensure laboratory instruments are calibrated on schedule against traceable reference standards, so that measurements and test results remain accurate and defensible.

2.Scope

Applies to fixed and portable measuring instruments used for testing, such as analytical balances, pH meters, thermometers, ovens, incubators and spectrophotometers. Pipettes are covered by the pipette calibration SOP.

Definitions

Reference standard
A certified weight, buffer solution or other traceable item used to check an instrument's accuracy.
Tolerance
The acceptable range of deviation from the true value for an instrument to remain in service.
Out of tolerance
A calibration result outside the acceptable range, requiring adjustment, service or removal from use.

3.Responsibilities

Calibration Technician
Performs scheduled calibrations, records results and applies calibration labels.
Lab Analyst
Checks the calibration label before using an instrument and reports suspected drift.
QA Manager
Maintains the calibration schedule, reviews certificates and approves instruments returning to service.
Laboratory Manager
Approves budget for outside calibration service and resolves recurring equipment issues.

RACI matrix

ActivityCalibration TechnicianLab AnalystQA ManagerLaboratory Manager
Maintain calibration schedule and due datesRIAI
Perform instrument calibrationR/AIC-
Approve out-of-tolerance instrument for service or retirementRIAC
Review calibration certificatesR-AI

R = Responsible, A = Accountable, C = Consulted, I = Informed

4.Materials and PPE

Materials, tools and systems

  • โ†’Certified reference weights, buffers and thermometers
  • โ†’Calibration log or CMMS asset register
  • โ†’Calibration due-date labels
  • โ†’Manufacturer calibration procedures
  • โ†’Out-of-service tags

Personal protective equipment

  • โ†’Disposable nitrile gloves
  • โ†’Lab coat

5.Procedure

  1. 5.1

    Review the calibration schedule

    Calibration Technician

    At the start of the week, check the calibration schedule or CMMS asset register for instruments due for calibration and pull them from the bench rotation.

  2. 5.2

    Gather reference standards

    Calibration Technician

    Select the certified reference weights, buffer solutions or thermometers required for the instrument, and confirm their certificates have not expired.

    Checkpoint: Reference standards used are within their own certification date.

  3. 5.3

    Inspect the instrument before testing

    Calibration Technician

    Check the instrument for physical damage, cleanliness and a stable, level, vibration-free location before starting the calibration procedure.

  4. 5.4

    Allow warm-up and stabilization

    Calibration Technician

    Power on the instrument and allow it to warm up and stabilize for the time stated in the manufacturer's instructions before taking readings.

  5. 5.5

    Perform the calibration procedure

    Calibration Technician

    Follow the manufacturer's calibration procedure for the instrument type, taking readings at the required points across its measuring range.

  6. 5.6

    Record readings and compare to tolerance

    Calibration Technician

    Record each reading against the reference value in the calibration log and compare the deviation to the instrument's stated tolerance.

    Checkpoint: Every reading is recorded before any adjustment is made to the instrument.

  7. 5.7

    Adjust or arrange service if out of tolerance

    Calibration Technician

    If a reading falls outside tolerance, adjust the instrument per the manufacturer's instructions and retest, or tag it out of service and arrange vendor repair.

    Warning: Do not return an out-of-tolerance instrument to use until it passes a repeat calibration.

  8. 5.8

    Apply the calibration label

    Calibration Technician

    Attach a label showing the calibration date, due date and technician initials to the instrument in a visible location.

  9. 5.9

    Update the asset register

    Calibration Technician

    Enter the calibration result, next due date and any parts replaced into the CMMS asset register for the instrument.

  10. 5.10

    Notify affected analysts

    Calibration Technician

    Tell the analysts who use the instrument that it is back in service, or that it remains out of service pending repair.

  11. 5.11

    Review certificates and trends

    QA Manager

    Review completed calibration certificates for the month, watching for instruments with repeated drift that may need replacement.

  12. 5.12

    File the calibration certificate

    QA Manager

    File the signed calibration certificate with the instrument's history file and retain it per the records retention schedule.

6.Quality checks

  • โ†’No instrument is used past its calibration due date.
  • โ†’Every calibration record shows the reference standard used and its certificate status.
  • โ†’Out-of-tolerance instruments are tagged out of service immediately.
  • โ†’Calibration certificates are filed and retrievable for every active instrument.

7.Records

  • โ†’Calibration log or CMMS asset register
  • โ†’Reference standard certificates
  • โ†’Calibration certificates
  • โ†’Out-of-service tags and repair records

8.KPIs

  • โ†’Percentage of instruments calibrated on schedule
  • โ†’Number of out-of-tolerance results per quarter
  • โ†’Average time an instrument is out of service for repair
  • โ†’Number of overdue calibrations

9.Common mistakes

  • โ†’Using an instrument past its calibration due date because the label was not checked.
  • โ†’Adjusting an instrument before recording the as-found reading.
  • โ†’Using an expired reference standard.
  • โ†’Not updating the asset register after a vendor repair.

10.Revision history

RevisionDateDescriptionReviewed by
1.0September 15, 2026Initial releaseIlia Pirozhenko

This is a template. Adapt it to your organization, equipment and local regulations before use.

Ask this SOP

Nobody opens a PDF in the middle of a task. They ask.

Add this SOP to Perfect Wiki and your team gets answers in the chat app they already use, with a link to the exact step. Ask from ChatGPT, Claude or Copilot too.

Perfect Wiki AIExample answer

The balance I need is out of tolerance today, can I still use it?

No. Tag it out of service, notify the calibration technician, and use a backup balance until it is recalibrated or repaired and passes a repeat calibration.Source: step 5.7 ยท Adjust or arrange service if out of tolerance
Ask your own question about this SOPโ€ฆSign up to keep asking
Word file vs Perfect Wiki

A downloaded SOP starts going out of date the day you save it.

Screen recorders like Scribe and Tango capture clicks. Perfect Wiki holds the whole procedure, including your recorded guides, and answers questions about it.

Word or PDFPerfect Wiki
Finding itDig through folders and email threadsAsk in Teams, Slack, kChat or Mattermost
Keeping it currentEmail a new version and hopeEdit once with AI, everyone sees the update
Everything in one placeText and imagesEmbed Scribe and Tango guides, SharePoint files and videos
Who can change itAnyone with the fileEditors you choose, everyone else reads
Common questions

Frequently asked questions

Didn't find what you're looking for? Contact our support โ†’

Your SOP library

Keep every SOP where your team can ask it.

Perfect Wiki is the knowledge base for Microsoft Teams, Slack, kChat and Mattermost. Store your SOPs, embed your Scribe and Tango guides, and let AI answer questions with a link to the right step.

No credit cardSetup in under 10 minutesCancel anytime