- Owner
- QA Manager
- Effective date
- September 15, 2026
- Review cycle
- Every 12 months
1.Purpose
To confirm each pipette delivers volumes within its stated accuracy and precision so that dilutions, dosing and test results made with it stay reliable.
2.Scope
Applies to all single-channel and multichannel air-displacement pipettes used for testing or sample preparation. General instrument calibration is covered by the equipment calibration SOP.
Definitions
- Gravimetric test
- A calibration method that weighs dispensed water on an analytical balance to calculate the actual volume delivered.
- Accuracy
- How close the average delivered volume is to the target volume.
- Precision (CV%)
- How consistent repeated deliveries are with each other, expressed as a coefficient of variation.
3.Responsibilities
- Lab Technician
- Submits pipettes for calibration on schedule and reports suspected inaccuracy between calibrations.
- Calibration Technician
- Performs the gravimetric test, records results and applies the pass or fail label.
- QA Manager
- Sets tolerance criteria, reviews failed results and approves pipettes returning to service.
RACI matrix
| Activity | Lab Technician | Calibration Technician | QA Manager |
|---|---|---|---|
| Submit pipette for scheduled calibration | R/A | I | - |
| Perform gravimetric calibration test | I | R/A | C |
| Approve pipette adjustment or service | I | R | A |
| Review calibration records and trends | - | R | A |
R = Responsible, A = Accountable, C = Consulted, I = Informed
4.Materials and PPE
Materials, tools and systems
- โAnalytical balance with calibration weights
- โDistilled or deionized water and thermometer
- โCalibration weights table for water density at temperature
- โManufacturer-approved pipette tips
- โCalibration log or CMMS asset register
- โPass or fail calibration labels
Personal protective equipment
- โDisposable nitrile gloves
5.Procedure
- 5.1
Check the calibration due date
Lab TechnicianCheck the label on the pipette or the CMMS asset register to confirm it is due, and remove it from the bench rotation for testing.
- 5.2
Record room conditions
Calibration TechnicianRecord the ambient temperature, humidity and barometric pressure in the calibration log, since these affect the water density calculation.
- 5.3
Inspect the pipette for damage
Calibration TechnicianCheck the shaft, tip cone, piston and seals for cracks, debris or wear, and clean the exterior before testing.
- 5.4
Perform a leak test
Calibration TechnicianAttach a tip, draw up liquid, and hold the pipette vertically for 20 to 30 seconds to check for a falling meniscus, which indicates a seal leak.
Warning: Do not proceed to gravimetric testing on a pipette that fails the leak test; tag it out of service.
- 5.5
Set the balance and tare
Calibration TechnicianPlace the weighing vessel on the analytical balance, close the draft shield, and tare the balance to zero before each dispense.
- 5.6
Dispense at minimum, mid and maximum volume
Calibration TechnicianDispense the target volume into the weighing vessel and record the weight for at least ten replicates at each of the minimum, mid and maximum volumes for that pipette.
Checkpoint: At least ten replicate weighings are recorded at each test volume before calculations begin.
- 5.7
Calculate accuracy and precision
Calibration TechnicianConvert each weight to volume using the water density at the recorded temperature, then calculate the mean volume, percent accuracy and percent CV for each test volume.
- 5.8
Compare results to tolerance
Calibration TechnicianCompare the calculated accuracy and precision against the manufacturer's stated tolerance for that pipette model and volume.
- 5.9
Adjust or tag out of service if failing
Calibration TechnicianIf results are out of tolerance, adjust the pipette per the manufacturer's instructions and retest, or tag it out of service and arrange vendor service.
- 5.10
Apply the pass or fail label
Calibration TechnicianAttach a label to the pipette showing the calibration date, due date, result and technician initials.
- 5.11
Log results and notify the owner
Calibration TechnicianEnter the results into the calibration log or CMMS and notify the pipette's regular user that it is back in service or remains out for repair.
- 5.12
Review failure trends
QA ManagerReview pipettes with repeated failures or adjustments across calibration cycles and decide whether replacement is more cost-effective than continued repair.
6.Quality checks
- โEvery calibration record shows at least ten replicates per tested volume.
- โRoom temperature is recorded and used in the volume calculation.
- โPipettes that fail the leak test are not gravimetrically tested until repaired.
- โPass or fail labels match the result recorded in the calibration log.
7.Records
- โCalibration log or CMMS asset register
- โGravimetric test data sheets
- โPass or fail calibration labels
- โVendor service records
8.KPIs
- โPercentage of pipettes calibrated on schedule
- โFirst-time pass rate on gravimetric testing
- โNumber of pipettes requiring adjustment per cycle
- โAverage pipette downtime for repair
9.Common mistakes
- โSkipping the leak test and going straight to gravimetric testing.
- โNot recording room temperature before calculating volumes.
- โTesting with tips that are not the manufacturer-approved brand for that pipette.
- โUsing fewer than the required number of replicates per volume.
10.Revision history
| Revision | Date | Description | Reviewed by |
|---|---|---|---|
| 1.0 | September 15, 2026 | Initial release | Ilia Pirozhenko |
This is a template. Adapt it to your organization, equipment and local regulations before use.