Pipette Calibration SOP Template

A ready-to-use SOP for gravimetric calibration of single and multichannel pipettes, covering leak testing, accuracy and precision checks, and pass or fail labeling.

FreeNo sign-upWord, PDF, Excel & CSV
Ilia PirozhenkoReviewed by Ilia Pirozhenko, Founder, Perfect WikiUpdated September 15, 202612 steps3 roles3 min read
Standard operating procedureSOP-LAB-006 ยท Rev 1.0
Owner
QA Manager
Effective date
September 15, 2026
Review cycle
Every 12 months

1.Purpose

To confirm each pipette delivers volumes within its stated accuracy and precision so that dilutions, dosing and test results made with it stay reliable.

2.Scope

Applies to all single-channel and multichannel air-displacement pipettes used for testing or sample preparation. General instrument calibration is covered by the equipment calibration SOP.

Definitions

Gravimetric test
A calibration method that weighs dispensed water on an analytical balance to calculate the actual volume delivered.
Accuracy
How close the average delivered volume is to the target volume.
Precision (CV%)
How consistent repeated deliveries are with each other, expressed as a coefficient of variation.

3.Responsibilities

Lab Technician
Submits pipettes for calibration on schedule and reports suspected inaccuracy between calibrations.
Calibration Technician
Performs the gravimetric test, records results and applies the pass or fail label.
QA Manager
Sets tolerance criteria, reviews failed results and approves pipettes returning to service.

RACI matrix

ActivityLab TechnicianCalibration TechnicianQA Manager
Submit pipette for scheduled calibrationR/AI-
Perform gravimetric calibration testIR/AC
Approve pipette adjustment or serviceIRA
Review calibration records and trends-RA

R = Responsible, A = Accountable, C = Consulted, I = Informed

4.Materials and PPE

Materials, tools and systems

  • โ†’Analytical balance with calibration weights
  • โ†’Distilled or deionized water and thermometer
  • โ†’Calibration weights table for water density at temperature
  • โ†’Manufacturer-approved pipette tips
  • โ†’Calibration log or CMMS asset register
  • โ†’Pass or fail calibration labels

Personal protective equipment

  • โ†’Disposable nitrile gloves

5.Procedure

  1. 5.1

    Check the calibration due date

    Lab Technician

    Check the label on the pipette or the CMMS asset register to confirm it is due, and remove it from the bench rotation for testing.

  2. 5.2

    Record room conditions

    Calibration Technician

    Record the ambient temperature, humidity and barometric pressure in the calibration log, since these affect the water density calculation.

  3. 5.3

    Inspect the pipette for damage

    Calibration Technician

    Check the shaft, tip cone, piston and seals for cracks, debris or wear, and clean the exterior before testing.

  4. 5.4

    Perform a leak test

    Calibration Technician

    Attach a tip, draw up liquid, and hold the pipette vertically for 20 to 30 seconds to check for a falling meniscus, which indicates a seal leak.

    Warning: Do not proceed to gravimetric testing on a pipette that fails the leak test; tag it out of service.

  5. 5.5

    Set the balance and tare

    Calibration Technician

    Place the weighing vessel on the analytical balance, close the draft shield, and tare the balance to zero before each dispense.

  6. 5.6

    Dispense at minimum, mid and maximum volume

    Calibration Technician

    Dispense the target volume into the weighing vessel and record the weight for at least ten replicates at each of the minimum, mid and maximum volumes for that pipette.

    Checkpoint: At least ten replicate weighings are recorded at each test volume before calculations begin.

  7. 5.7

    Calculate accuracy and precision

    Calibration Technician

    Convert each weight to volume using the water density at the recorded temperature, then calculate the mean volume, percent accuracy and percent CV for each test volume.

  8. 5.8

    Compare results to tolerance

    Calibration Technician

    Compare the calculated accuracy and precision against the manufacturer's stated tolerance for that pipette model and volume.

  9. 5.9

    Adjust or tag out of service if failing

    Calibration Technician

    If results are out of tolerance, adjust the pipette per the manufacturer's instructions and retest, or tag it out of service and arrange vendor service.

  10. 5.10

    Apply the pass or fail label

    Calibration Technician

    Attach a label to the pipette showing the calibration date, due date, result and technician initials.

  11. 5.11

    Log results and notify the owner

    Calibration Technician

    Enter the results into the calibration log or CMMS and notify the pipette's regular user that it is back in service or remains out for repair.

  12. 5.12

    Review failure trends

    QA Manager

    Review pipettes with repeated failures or adjustments across calibration cycles and decide whether replacement is more cost-effective than continued repair.

6.Quality checks

  • โ†’Every calibration record shows at least ten replicates per tested volume.
  • โ†’Room temperature is recorded and used in the volume calculation.
  • โ†’Pipettes that fail the leak test are not gravimetrically tested until repaired.
  • โ†’Pass or fail labels match the result recorded in the calibration log.

7.Records

  • โ†’Calibration log or CMMS asset register
  • โ†’Gravimetric test data sheets
  • โ†’Pass or fail calibration labels
  • โ†’Vendor service records

8.KPIs

  • โ†’Percentage of pipettes calibrated on schedule
  • โ†’First-time pass rate on gravimetric testing
  • โ†’Number of pipettes requiring adjustment per cycle
  • โ†’Average pipette downtime for repair

9.Common mistakes

  • โ†’Skipping the leak test and going straight to gravimetric testing.
  • โ†’Not recording room temperature before calculating volumes.
  • โ†’Testing with tips that are not the manufacturer-approved brand for that pipette.
  • โ†’Using fewer than the required number of replicates per volume.

10.Revision history

RevisionDateDescriptionReviewed by
1.0September 15, 2026Initial releaseIlia Pirozhenko

This is a template. Adapt it to your organization, equipment and local regulations before use.

Ask this SOP

Nobody opens a PDF in the middle of a task. They ask.

Add this SOP to Perfect Wiki and your team gets answers in the chat app they already use, with a link to the exact step. Ask from ChatGPT, Claude or Copilot too.

Perfect Wiki AIExample answer

My pipette failed the leak test, can I still send it for gravimetric calibration today?

No. Tag it out of service and set it aside for repair. A pipette that fails the leak test will also fail or give unreliable gravimetric results, so it needs a seal repair first.Source: step 5.4 ยท Perform a leak test
Ask your own question about this SOPโ€ฆSign up to keep asking
Word file vs Perfect Wiki

A downloaded SOP starts going out of date the day you save it.

Screen recorders like Scribe and Tango capture clicks. Perfect Wiki holds the whole procedure, including your recorded guides, and answers questions about it.

Word or PDFPerfect Wiki
Finding itDig through folders and email threadsAsk in Teams, Slack, kChat or Mattermost
Keeping it currentEmail a new version and hopeEdit once with AI, everyone sees the update
Everything in one placeText and imagesEmbed Scribe and Tango guides, SharePoint files and videos
Who can change itAnyone with the fileEditors you choose, everyone else reads
Common questions

Frequently asked questions

Didn't find what you're looking for? Contact our support โ†’

Your SOP library

Keep every SOP where your team can ask it.

Perfect Wiki is the knowledge base for Microsoft Teams, Slack, kChat and Mattermost. Store your SOPs, embed your Scribe and Tango guides, and let AI answer questions with a link to the right step.

No credit cardSetup in under 10 minutesCancel anytime